WaterSam WS 312 is a space-saving sampler for all standard indoor and outdoor applications and together with its built-in cooling and heating features, samples are able to store reliably at the require desired temperature.
Features:
- Weather-proof housing and roof made of durable stainless steel; upper panel door included
- Three separate technical compartments in the upper section of the sampler to protect the pump, electronics and refrigeration unit
- Freely adjustable and highly accurate temperature control for the sample storage chamber, for ambient temperatures from -25°C to +42°C
- High-performance vacuum sampling system – alternatively available with any other WaterSam sampling system
- Borosilicate glass metering vessel protected in sample storage chamber to minimize sample falsication that could stem from temperature extremes
- Hose inlet on the left and right side; optionally from below
- Available with XY Distributor for direct filling of numerous bottle sets without cross-contamination
- Simple upgrade to monitoring station can be made at anytime
- Wide range of optional equipment: housing materials, pumps, valves, etc.
Direct Bottle Filling with Numerous Possibilities
WaterSam XY Distributor system has been in use successfully for nearly 20 years and with its wide array of possibilities, reliable operation and minimal follow-up costs have made it the distributor of choice among our customers. The two-axis system makes full use of the entire sample storage chamber by using coordinates to travel and locate bottle positions.
Technical Specifications
General | Stationary sampler for indoor and outdoor applications |
Norms | CE, compliant with ISO 5667 |
Dimensions | 1,020 x 590 x 590 mm (HxWxD) Width including mounting rails: 720 mm |
Weight | Approx. 70-80 kg, depending on Configuration |
Power Requirements | 230 V AC (optional: 110 V AC), 50 Hz (optional: 60 Hz), Main fuse 16 A Approx 750 Watts Max include High Performance Cooling or Heating |
Ambient Temperature | -25 °C to +42 °C |
Sampling System | VAC vacuum pump system for time-, volume-, event-proportional sampling Optional: VAR-B, VAR-E, VAR-C or peristaltic pump for flow-proportional sampling (variable volumes) |
Sampling Volume | 12 – 200 ml (multiple-shot sampling possible) Optional: up to 2000 ml (without distributor) |
Sample Vessel | DURAN 50 borosilicate glass; dishwasher-safe, resistant to acid, alkali, and temperature fluctuations |
Pump Performance | Max. 8 m lift height; 0.5 m/s up to 6 m lift heights; 14 l/min free flow Optional: – high-performance pump (19 l/min) – VacuPress for lift heights up to 30 m |
Intake Hose | 12 mm ID PVC |
Wetted Parts | Borosilicate glass, PE, PVC, stainless steel 304/316Ti, silicone (optional: alternative materials as required) |
Housing Material | Stainless steel 304 Optional: – Stainless steel 316Ti – Powder coating (RAL colors) – Plastic (UV resistant) |
Insulation | 40 mm, cold-bridge free; not foamed |
Environmental Consideration | Construction facilitates easy separation of materials for proper recycling and disposal |
Partitioning | Three separate technical compartments in top dry section for electrical/electronics, refrigeration unit and other components |
Placement of Sampling System | Metering vessel in temperature-controlled sample chamber; protected from heat and frost |
Installation | Easy floor mounting and firm footing with sturdy base rails |
Temperature Control | PT 100 3-point thermostat |
Refrigeration | Compressor cooling; 230 V AC, 150 W; CFC-free refrigerant R134a; adjustable automatic defrosting (interval, time, duration, max. temperature) Optional for corrosive environment, e.g. H2S: evaporator plate separate from sample storage chamber, clean external air drawn to cabinet via ventilation flange |
Heating | Electric heating unit in stainless steel sleeve; 230 V AC, 350 W (optionally 24 V DC) |
Sample Storage | Temperature adjustable; pre-set to 3 °C |
Distributor (Optional) | |
Distributor | XY Distributor: two-axis coordinate system for direct depositing of discrete samples with no cross-contamination |
Bottle Synchronization | Automatic |