The MS2000-S is a highly sensitive on-line total trihalomethane analyser which helps water
treatment operators optimise their process to meet local regulations.
The MS2000 utilises a contactless measurement technique that greatly reduces maintenance requirements. Furthermore, the instrument does not require reagents, which keeps the ongoing running costs to a minimum.

Features:
- NO REAGENTS or GASES: Low Running Costs
- The most cost-effective solution
- High sensitivity: ideal for potable water
- No sensor contact with water
- Low maintenance, no sensor cleaning
- Versatile communication protocols
- Hourly Sampling with Instant Results
Applications:
- Monitor THMs post-chlorination
- Analyse THMs in the distribution network
- Swimming pools and water parks
- Detect RO membrane breakthrough
- THMs in food & beverage production
- Monitoring THMs in large establishments
| Detection Range | 5 ~ 200 ppb (ug/L) as Total THMs |
| Accuracy | +/- 10% * |
| Repeatability | +/- 2% |
| Measuring Frequency | 60 Mins |
| Analogue Output | 1 x 4 ~ 20mA, Max Load @ 750 Ohms |
| Relay Output | 2 x Alarm 1 x Fault NO and NC Contact, Max: 250VAC or 30VDC, 5A |
| Interface | TouchScreen Panel PC |
| Power Supply [Power Consumption] | 90 ~ 240VAC, 50/60Hz [Max: 25 Watts] |
| Communication Protocols | Modbus (Optional) / Profibus / HART/ Foundation Fieldbus on request |
| Data Storage | uSD Card / Windows PC |
| Dimension [Weight] | 450 x 680 x 200mm (L x H x D) [30 kg] |
| Enclosure IP | IP65 Powder Coated Stainless Steel |
| Sample Flow Rate [Temperature] | 2L/min [1 ~ 50 deg C] |