LAR QUICKTOCpurity – Pure Water, Condensate Return, Drinking Water and Treated Effluent Samples

LAR QuickTOCpurity Analyser is the most reliable measurement system for the Pure Water, Condensate, Drinking Water and Treated Effluent Application

Due to high temperature combustion at 1200 deg C, this model is excellent in fully Oxidizing all Carbon Bonds.

Almost all type of sample can be measured accurately by this unit. Due to combustion at 1200 deg C, NO catalyst is needed.

LAR’s QuickTOCpurity is the ultimate online TOC analyser for the determination of total carbon (TC), total organic carbon (TOC), total inorganic carbon (TIC) and dissolved organic carbon (DOC) for Pure Water, Condensate, Drinking Water and Treated Effluent Application

LAR QuickTOCpurity Analyser with Sample Injection to Furnace

Features:

  • Continuous TC, TOC, TIC monitoring
  • Easy calibration and validation (QuickCalibration)
  • Highest combustion temperature (1,200°C) available on the market
  • Catalyst-free technique
  • Closed system to prevent contaminations
  • No memory effects
  • Low operational and maintenance costs
  • According to DIN EN 1484:1997-08, ISO 8245:1999-03 and EPA 415.1

Applications:

  • Treated Effluent water
  • Discharge control
  • Treated Industrial waste water
  • River Water
  • Drinking Water
  • Ultra Pure Water
  • Pure Water
  • Condensate Return

Technical Specifications

Measuring PrincipleHigh Temperature Combustion at 1200 deg C without Catalyst and
measure Carbon Dioxide (CO2 ) Produces with NDIR Carbon Dioxide (CO2) Detector
Measuring ParameterTotal Carbon (TC), Total Inorganic Carbon, Total Organic Carbon
Measuring Range
Accuracy and Reproducibility+/- 2%
Analyser SamplingPeristatic Pump with High Reliability Solenoid Valves
DisplayHigh res. 10.4-inch backlit LCD touch screen
Standard FeaturesAuto Start Function, Self Explanatory Software
Outputs0/4 – 20mA for Individual Measuring Stream
8 x Dry Contact Relays
InterfaceRS232
USB 2.0
Analysis StartProgrammable Time Interval
External Digital Signal [Remote Start]
Standard FeaturesAuto Start Function
Self Explanatory Software
Power Supply
[Power Consumption]
100VAC or 230VAC, 50/60 Hz,
[Max: 1.2 kW | Average: 750 Watts]
Standard Enclosure
[Dimension]
Splash Proof Steel Enclosure (IP54), Powder Coated
[1 Sample Stream: 600 x 1,062 x 608 mm (W x H x D)]
[2 – 6 Sample Streams: 820 x 1,062 x 608 mm (W x H x D)]
Additional OptionUpgrade to 6 Streams in 1 Analyser
Upgrade with COD* based on Correlation from TOC and / or TNb** Measurement in 1 Analyser
Upgrade with Remote Access via TCP/IP
Enclosure OptionIP65 (NEMA 4X) Stainless Steel Enclosure
Ex Proof Purge Stainless Steel Enclosure for ATEX or IECex Area
Note* Chemical Oxygen Demand, ** Total Nitrogen Bound
Option with Ex-Proof Purge Enclosure for Install Unit in ATEX or IECex Area

A I ANALYTICAL PTE LTD | Address 30, Kallang Place, #06-01, Singapore 339159 | Tel: +65 6459 0785 | Email: info@ai-analytical.com.sg